BESSY MySpot 05-11 January 2026

This post provides a journey through the entire experiment, including sample preparation, experimental design and data evaluation.

Coral sample preparation: 200 micron thick embedded coral slices

1) Sample labeling
→ every sample was assigned with numbers: 500 - 505
→ every sample side: a, b, c, etc.
→ p = Pocilopora
→ s = Stylophora

1a) Samples measured (highlighted in yellow) with XRD, XRF, absorption –> RADIATION DAMAGE (KATREIN):

Species Sample nomenclature      
Pocillopora p500a (175 mu) p500b (340 mu) p500c (376 mu) ——-
Pocillopora p501a (306 mu) p501b (298 mu) p501c (180 mu) ——-
Pocillopora p502a (176 mu) p502b p502c (160 mu) ——-
Stylophora Pistillata s503a (330 mu) s503b (345 mu) s503c (204 mu) s503d (333 mu)
Stylophora Pistillata s504a (145 mu) s504b (216 mu) s504c (343 mu) s504d (333 mu)
Stylophora Pistiallata s505a (314 mu) 505b (256 mu) s505c (299 mu) s505d (167 mu)

1b) Samples measured with XRD, XRF, absorption –> FLOW / NO FLOW (ISABELA):

Species Sample nomenclature
Pocillopora grandis pg31b2
Pocillopora grandis pg07b2
Pocillopora grandis pg11t2
Pocillopora grandis pg31t2

2) Raw sample cutting
→ Untreated corals were cut into 1-1.5 mm thick slices:
1. Embed end of sample in thin layer of pmma, which then can be stuck to the transparent rectangle that fits in the suction plate of the bandsaw, to better control quality and thickness of slices of raw coral
Bandsaw (Name, company,location):

3) Embedding
Each sample was potted in methyl-methacrylate (Technovit 4071, Kulzer GmbH, Germany). Firmly secured at the potted ends, samples were then sectioned cross-sectionally into ca. mm-thick sections. Each section was embedded individually (Technovit 4071, Kulzer GmbH, Germany), so as to infiltrate otherwise closed internal pores.

→ Embed these slices in pmma, filling the mold to the top.
→ Side A = bottom of embedding mold = side of puck with sample near surface
→ Side B = top of embedding mold = far from sample
→ End up with a puck, where thin slice of sample is at one end, and the rest is pmma.

4a) Sample polishing short for paper
Embedded pucks containing the sample section were then ground to expose the coral skeleton material and polished (EXAKT 400 CS, EXAKT Advanced Technologies GmbH, Germany) on both sides, to produce 200-400 µm-thick highly polished plane-parallel slices. Specifically, blocks were first roughly ground down to ca. 500 um thick (800 grit), making sure that the coral skeleton was exposed on both sides of the puck. Then, each side was polished using the following protocol: 1. 1 min at 1000 grit with water 2. 1 min at 1200 grit with water 3. 3 minutes at 2500 grit with water 4. 3 minutes at 4000 grit with Na2CO3 All polishing steps were done with xx g weight. The final step at 4000 grit was done with Na2CO3 to prevent surface ACC dissolution (YUT Gong et al., PNAS 2012). Samples were then wiped (Kimtech wipe with Na2CO3), dried (pressurized air), and stored in a clean membrane box.

4b) Sample polishing in detail: Polish side A
→ Can work in batches of ~6 samples.
→ Polish side A (removes 0.1 mm):
1. Manually sand side B with 800 grit paper, just to make sure surface is clean and smooth and will stick well with tape. Wipe dust off with wet tissue and dry surface with compressed air duster.
2. Stick samples (side B stuck, side A exposed) onto transparent rectangle
3. Polishing standard procedure (Removes 0.1mm total):
- 1000 grit, 1 small weight, ~2 minutes (or more if needed) to expose surface
- 1 min @ 1000 grit, 1x small weight
- 1 min @ 1200 grit, 1x small weight
- 3 min @ 2500 grit, 1x small weight
- 3 min @ 4000 grit, 1x small weight

4c) Polish side B
→removes 0.1 mm, Plastic rectangle thickness + 1 layer of tape = 2.2 mm → target thickness = 2.2 + 0.3 = 2.5mm
→ Use bandsaw to cut off excess pmma from side B, leaving pucks of ~1.5mm thick (with the coral slice inside).
→ Grind side B (800 grit, 2x small weitht + large weight) until sample is 0.3 mm thick (0.3 + 2.2 = 2.5 sample plastic tape unit). In polisher machine, can use the meter knob to set thickness at which polishing will stop. Make sure to zero without sample.
→ Stick samples (side A stuck, side B exposed) onto transparent rectangle: To allow eventual detachment of sample without damaging delicate slice, stick it with thin slivers of tape only on two sides of the puck where there is just pmma.
→ This will be slow, goes faster if use little water. Can start with 320 grit and move to 800 once approaching target thickness.
→ Then do standard polishing as above
→ Remove samples from plastic rectangle. Use fine dentistry tools to slowly lift tape off of plastic rectangle.
→ Wipe slice with kimwipe and solution of filtered 22g/L Na2CO3 (in fridge in front of the door of the room where Keyence is, in glass bottle with blue cap, labelled…you will see it).

Note: Not further polished with diamond/silica/alumina paste because in test sample this standard polishing was enough to reach < 1um roughness on side B (last polished surface).

In the end, the polishing machine is not that precise that I cut the samples with the band saw as thin as possible. If I chose to cut the sample too thin, the band saw makes funny things and the sample just breaks. After cutting I polished with 800 grit polishing paper and polishing machine till a thickness of approximately < 1 mmm (determined with blank eyes). Then increasing stepwise the grit of the polishing paper (see polish side A) till I am happy with the surface including thickness. At the end I measured the thickness precisely and noted the size in the table above.

5) Before beamtime measurements:

Samples were measured with light microscopy

6) At the beamtime:

Participants: Katrein Sauer, Paul Zaslansky, Isabela Vitienes, Shreya Ray

Samples were measured within a diaphrame samdwiched within two yellow Kapton films / polyimide films (Kapton-Folie / Polyimid-Folie, Imid-group: -CO-NH-CO- –> electric isolator, thermal stable and chemical- and radiation resistant).

Energy: 17 keV
beamsize: 50 microns\

Plan:

  1. Mapping the entire samples with a stepsize of 100 x 100 µm
  2. choose two ROIs with a stepsize of 50 x 50 points (=highest possible resolution defined by beamsize) with 1s or 2s exposure time
  3. Irradiate several choosen spots over and over again, and collect XRD patterns (Debye rings) over time

Close the Hutch:

  • press green button on the wall
  • close door
  • rotate key right beside the door and take it with your
  • put key into beamline lights which shows green light if everything works well (MySpot (NOT BAMline!)) and rotate key 1/4 to the right
  • press leverage (Hebel) up and release. It will beep and every diode in that line should light up green now

Open the Hutch:

  • rotate key from 0–>1 and take it with you to the pace right beside the door
  • enter key and bring it to position 1
  • press green button
  • door will make a noise, you can open it
**Helpful commands:**    -   beamstop_out    - PAUSE/STOP: \
P [=pause] \
r [=resume]\
Strg + C [=der totale Abbruch]
- curs [= to go anywhere within ascan with motor] \
 m [= move motor to that position]
 q [= quit]
- wm kox [= where motor (position) kox]
- wm mz [= where motor (postition) mz]

7) Additional measurements after beamtime:

  1. Sample s504d for XRF to look into Magnesium (PTB beamline (Adrian))
  2. Leica light microscopy
  3. SEM-BEI
  4. Laminography (Isabela)

8) Evaluation

Transmission / Absorption (diode1):   - diode1 = absolute absorption, good calibration, gives photons / s   - cyber = transmission (not as good calibrated as beam + has to go to through air and through the sample --> absortion of both. Therefore: transmission = air / sample)


XRF FLUORESCENCE:

Measured elements as part of coral samples:

  calcium (Ca) zinc (Zn) strontium (Sr)
channels [ROI] 338-378 [1] 823-863 [2] 1340-1415 [3]
Energy [keV] 3.69 8.638 14.165
Color (Fiji) orange hot magenta hot thallium

Measured elements as part of background / beamline setup / contamination:

  Iron (Fe) Arsen (As) Lead (Pb) Cupper (Cu) Beam
channel (maximum) 625 1029 ??? 1029 ??? 1230 1652
Energy [keV] 6.35 10.53 10.55 12.6 17

The arsen (K_alpha) and lead (L-line) energy are both included in one peack as this peak is broad enough to carry both. Therefore, it is not clear which element it really is exactly but most likely LEAD!

From measurement set68 till set74: Here is a mistake in naming: Instead of sample p501c, it actually is the sample s503a and must therefore be set68_s503a,…,set74_s503a, and also set76_s503a,…,set89_s503a!!! The same is true for XRD!

XRD X-Ray diffraction (WAXS):

Energy: 17 keV
beamsize: 50 microns
SSD:
Detector: Eiger 9M
Diaphrame size:
mz = 0 - 25 (window inner size = 24 mm)
koi = -20 -0 - +25 (window inner size = 36 mm)
–> Richtung -20: fährt man zur Wand

All .h5-files were transformed to .tiff for further evaluation using Fiji.

From measurement set68 till set74: Here is a mistake in naming: Instead of sample p501c, it actually is the sample s503a and must therefore be set68_s503a,…,set74_s503a and also set76_s503a,…,set89_s503a!!! The same is true for XRF!

**9) XRD Integration from Ivo**

Ivo only took one corundum for all the samples: NEEDS TO TAKE ALL CORUNDUMS LEFT AND RIGHT OF THE Samples

INTEGRATED “CALCITE” IMAGES ACTUALLY BELONG TO AN ARAGONITE PEAK!!!

We see already huge and very few single calcite peaks (~ 1µm-range?!)

Written on December 16, 2025